Qualification Limits

Specific qualification tests are required to demonstrate proper performance of systems for safety related applications. The Meggitt Safety Systems CAMS systems have been subjected to the complete compliment of tests for which the primary governing specifications include:

    Seismic IEEE 344-1975
    Environmental IEEE 323-1974
    EMI EPRI TR-102323

The "As Tested" environmental conditions envelope most site specific requirements.

Qualification Limits

Parameter

Remote Assembly

Analyzer Assembly

Loop Components, Qualified, 1)

Temperature, ° F

40 – 120

60-120

60 to 306 (2)

Pressure, ATM Absolute

1

1

0.7 to 5.0

Humidity, R.H.

0 – 95

20 – 90

20 to 100

Radiation, RADs, TID

1x10 4

1.0x10 6
GAMMA

2.0 x 10 8,
GAMMA and BETA

Seismic,
Peak G’s

9, Horizontal
9, Vertical

4.7, Horizontal
5, Vertical

4.7, Horizontal
5, Vertical

  1. Components in direct contact with the sample.
  2. Post accident, containment atmosphere includes super-heated steam. By terminating the heat trace on the inlet sample line upstream of the analyzer approximately 8 feet, the super-heat will be lost as the sample temperature cools down to the dew point, a temperature within the MSSI qualification limit.

 

 

©2005-2010 Meggitt Safety Systems